Section 5 Alarm Descriptions
Setting Alarms
Limits are selected in the SETUP Menu by choosing a value using the UP/DOWN or
CURVED ARROW knobs, or the UP or DOWN Navigation buttons.
Sensitivity and Duration are two general terms used in the alarm menus:
Sensitivity
Sensitivity is set by Consecutive Errored Samples (CES). When setting amplitude
limits, a noise spike can exceed the limit while the video amplitude can be within
the limit. With the CES set to a low number, a spike is detected and an alarm is
displayed. Set the CES to a higher number to ignore the fast spike. Each CES
occurs at 37ns intervals for SD. Use this as a guideline to select the appropriate
CES value.
Not all alarms have CES associated with them; in such cases, use the duration to
increase or decrease the general sensitivity.
Duration
This value is used to determine how long an error must persist before it is reported.
Set the duration to 0 to cause an error to be displayed as soon as the CES value is
met.