VTM-2000 Installation and Operation Handbook
5-1
Section 5 ♦ Alarm Descriptions
Setting Alarms
Limits are selected in the SETUP Menu by choosing a value using the UP/DOWN or
CURVED ARROW knobs, or the UP or DOWN Navigation buttons.
Sensitivity and Duration are two general terms used in the alarm menus:
• Sensitivity
Sensitivity is set by Consecutive Errored Samples (CES). When setting amplitude
limits, a noise spike can exceed the limit while the video amplitude can be within
the limit. With the CES set to a low number, a spike is detected and an alarm is
displayed. Set the CES to a higher number to ignore the fast spike. Each CES
occurs at 37ns intervals for SD. Use this as a guideline to select the appropriate
CES value.
Not all alarms have CES associated with them; in such cases, use the duration to
increase or decrease the general sensitivity.
• Duration
This value is used to determine how long an error must persist before it is reported.
Set the duration to 0 to cause an error to be displayed as soon as the CES value is
met.